PLZ-4W_General Specifications

ModelPLZ164WPLZ334WPLZ1004WPLZ164WAPLZ664WA
Input voltage range100 VAC – 240 VAC (90 VAC – 250 VAC) Single phase, continuous100 VAC – 120 VAC/200 VAC – 240 VAC (90 VAC – 132 VAC/180 VAC – 250 VAC) Single phase
Input frequency range47 Hz to 63 Hz
Power80 VA max90 VAmax160 VAmax300 VAmax1500 VAmax *1
Inrush current45 A80 A
Operating temperature range0 °C to 40 °C (32 °F to 104 °F)
Operating humidity range20 % – 85 % RH (without condensation)
Storage temperature range-25°C to 70°C (-13 °F to 158 °F)
Storage humidity range90 % RH or less (without condensation)
Isolation voltage± 500 V
Insulation resistanceBetween primary and input terminals500 VDC, 30 MΩ or more (ambient humidity of 70 % RH or less)
Between primary and chassis500 VDC, 30 MΩ or more (ambient humidity of 70 % RH or less)
Withstanding voltageBetween primary and input terminalsNo abnormalities at 1500 VAC for 1 minute.
Between primary and chassisNo abnormalities at 1500 VAC for 1 minute.
WeightApprox. 7 kg (15.43 lb)Approx. 8 kg (17.64 lb)Approx. 15 kg (33.1 lb.)Approx. 7.5 kg (16.53 lb)Approx. 16 kg (35.27 lb)
Battery backupBacks up setup information
AccessoriesPower cord × 1 pc. (with plug, Length: approx. 2.4 m), Load input terminal cover × 1 piece (2 lock plates provided), Set of screws for the load input terminal × 2 sets (bolts, nuts, and spring washers), Setup guide × 1 piece(Japanese, English), Quick Reference(Japanese:1 piece, Englis:1 piece), CD-ROM × 1 piece *2
Electromagnetic compatibility (EMC) *3, *4Conforms to the requirements of the following standard.
EMC Directive 2014/30/EU
EN 61326-1 (Class A *7), EN 55011 (Class A *7, Group 1 *8), EN 61000-3-2 *9 ,
EN 61000-3-3 *9
Safety *3Conforms to the requirements of the following standard.
Low Voltage Directive 2014/35/EU *4 EN 61010-1 (Class I *5, Pollution degree 2 *6)

*1. 900 VA when input voltage is 100 VAC. *2. Contains Application & Samples, User’s manual, the Communication Interface Manual and KI-VISA. *3. Does not apply to specially ordered or modified PLZ-4Ws. *4. Limited to products that have the CE mark on their panels. Not be in compliance with EMC limits unless the ferritecore is attached on the cable for connection of J1 connector. *5. This is a Class I equipment. Be sure to ground this product’s protective conductor terminal. The safety of this productis only guaranteed when the product is properly grounded. *6. Pollution is addition of foreign matter (solid, liquid or gaseous) that may produce a reduction of dielectric strength orsurface resistivity. Pollution Degree 2 assumes that only non-conductive pollution will occur except for an occasionaltemporary conductivity caused by condensation. *7. This is a Class A equipment. This product is intended for use in an industrial environment. This product may causeinterference if used in residential areas. Such use must be avoided unless the user takes special measures to reduceelectromagnetic emissions to prevent interference to the reception of radio and television broadcasts. *8. This is a Group 1 equipment. This product does not generate and/ or use intentionally radio-frequency energy, in theform of electromagnetic radiation, inductive and/or capacitive coupling, for the treatment of material or inspection/analysis purpose. *9. Excluding the PLZ664WA.