TOS9311_General Specifications

EnvironmentInstallation locationIndoors, 2000 m or less, Pollution Degree 2
Spec guaranteed rangeTemperature5 °C to 35 °C (41 °F to 95 °F)
Humidity20 %rh to 80 %rh (no condensation)
Operating rangeTemperature0 °C to 40 °C (32 °F to 104 °F)
Humidity20 %rh to 80 %rh (no condensation)
Storage rangeTemperature-20 °C to 70 °C (-4 °F to 158 °F)
Humidity90 %rh or less (no condensation)
Power supplyNominal voltage range (allowable voltage range)100 Vac to 120 V, 200 V to 240 V (90 Vac to 132 V, 170 V to 250 V), no switching required
Power consumptionNo load (READY state)100 VA or less
Rated load800 VA max.
Allowable frequency range
47 Hz to 63 Hz
Insulation resistance (between AC LINE and chassis)30 MΩ or more (500 Vdc)
Withstanding voltage (between AC LINE and chassis)1500 Vac, 1 minute, 20 mA or less
Ground bond25 Aac, 0.1 Ω or less
Electromagnetic compatibility *2 *3Complies with the requirements of the following directive and standard.
EMC Directive 2014/30/EU EN 61326-1 (Class A), EN 55011 (Class A, Group 1)
EN 61000-3-2, EN 61000-3-3
Applicable under the following conditions: The maximum length of all cabling and wiring connected to the product must be less than 2.5 m. Shielded cables are being used when using the SIGNAL I/O. Use the included high voltage test leads. Electrical discharges are applied only to the EUT.
SafetyComplies with the requirements of the following directive and standard.
Low Voltage Directive 2014/35/EU *1
EN 61010-1 (Class I, Pollution Degree 2), EN61010-2-030
Dimensions(MAX)/Weight430(16.93) (440(17.32)) W × 174.2(6.86) (195(7.68)) H × 500(19.69) (540(21.26)) D
mm(inches) / Approx. 27 kg (59.5 lb.)
AccessoriesPower cord (1 pc.), High-voltage test lead (1 pair),
SIGNAL I/O plug (1 set) Assembly type D-sub plug unit,
High-voltage warning sticker (1 pc.), Heavy object warning label (1 pc.), Cable tie (1 pc.), Getting Started Guide (1 copy), Safety Information (1 copy), China RoHS sheet (1 sheet)

*1 Only on models that have CE/UKCA marking on the panel.