Does TOS7200 have a function to output the measured resistance values as an analog voltage?

Yes, TOS7200 is equipped with an ANALOG OUT terminal that outputs a logarithmically compressed voltage from 0 to 4 V based on the resistance measurement Rx:

  • 1 MΩ: 0.30 V
  • 10 MΩ: 1.04 V
  • 100 MΩ: 2.00 V
  • 1000 MΩ: 3.00 V
  • 10,000 MΩ or higher: 4.00 V

*Output impedance: 1 kΩ