Firmware Version 2.14
2025-1-10
IFC2.14.0363/IOC2.00.0354/FPGA1.04.0007
Improvement
Integrated TOS9300 / TOS9301 / TOS9302 / TOS9303 / TOS9303LC / TOS9301PD / TOS9311 models.
Updated
Changed ACW / DCW / IR / PD modes to add graph functionality.
Updated
Changed Key Lock to allow password setting.
Updated
Changed to allow users to set a custom calibration date.
Firmware Version 2.15
2025-4-25
IFC2.15.0366/IOC2.00.0354/FPGA1.04.0007
Resolved issues
Fixed the problem where the ACW mode status (normally green) to be displayed in red at power-on in rare cases.
Firmware Version 2.16
2025-9-08
IFC2.16.0369/IOC2.00.0361/FPGA1.04.0007
Resolved issues
Fixed the problem that was preventing the INITiate command from being executed while the DANGER LED was flashing.
Improvement
Integrated NRTL models.
Firmware Version 2.17
2026-5-28
IFC2.17.0372/IOC2.00.0373/FPGA1.04.0007
Resolved issues
Fixed the issue where a malfunction occurred in the IR test during Auto mode on TOS9300 models when the Judge Delay was changed. brbr*Only for TOS9300 models.
Resolved issues
Fixed the issue where a QUERY error could occasionally occur when using the MEAS? command.
Resolved issues
Fixed the issue where the U-FAIL detection response in the DCW test became unstable. brbr*Occurred in Ver. 2.14 to Ver. 2.16.
Resolved issues
Fixed the issue where the Current Real value in the Result screen of the ACW test was displayed as the Current value. brbr*No issue with commands or CSV files.
Resolved issues
Fixed minor issues related to graph drawing.
Improvement
Improved the response when switching to the high-sensitivity range during IR tests.